Evaluating and defining the sharpness of intraocular lenses: microedge structure of commercially available
Liliana Werner1, Matthias Müller, Manfred Tetz
1Berlin Eye Research Institute, Berlin, Germany. werner.liliana@gmail.com
Purpose:
To evaluate the microstructure of the edges of currently available square-edged hydrophobic intraocular lenses (IOLs) in terms of their deviation from an ideal square.
Setting:
Berlin Eye Research Institute, Berlin, Germany.
Methods:
Sixteen designs of hydrophobic acrylic or silicone IOLs were studied. For each design, a +20.0 diopter (D) IOL and a +0.0 D IOL (or the lowest available plus dioptric power) were evaluated. The IOL edge was imaged under high-magnification scanning electron microscopy using a standardized technique. The area above the lateral-posterior edge, representing the deviation from a perfect square, was measured in square microns using reference circles of 40 microm and 60 microm of radius and the AutoCAD LT 2000 system (Autodesk). The IOLs were compared with an experimental square-edged poly(methyl methacrylate) (PMMA) IOL (reference IOL) with an edge design that effectively stopped lens epithelial cell growth in culture in a preliminary study. Two round-edged silicone IOLs were used as controls.
Results:
The hydrophobic IOLs used, labeled as square-edged IOLs, had an area of deviation from a perfect square ranging from 4.8 to 338.4 microm(2) (40 microm radius reference circle) and from 0.2 to 524.4 microm(2) (60 microm radius circle). The deviation area for the square-edged PMMA IOL was 34.0 microm(2) with a 40 microm radius circle and 37.5 microm(2) with a 60 microm radius circle. The respective values for the +20.0 D control silicone IOL were 729.3 microm(2) and 1525.3 microm(2) and for the +0.0 D control silicone IOL, 727.3 microm(2) and 1512.7 microm(2). Seven silicone IOLs of 5 designs had area values that were close to those of the reference square-edged PMMA IOL. Several differences in edge finishing between the IOLs analyzed were also observed.
Conclusions:
There was a large variation in the deviation area from a perfect square as well as in the edge finishing, not only between different IOL designs but also between different powers of the same design. Clinically, factors such as the shrink-wrapping of the IOL by the capsule may even out or modify the influence of these variations in terms of preventing posterior capsule opacification.
More Related Videos
Related Concept Videos
Punnett Squares
Root Mean Square
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...
Chi-square Analysis
The chi-square test was developed by Pearson in 1990.
The first step of performing a Chi-square analysis is to establish a null hypothesis, which assumes that there is no real...
Crystal Field Theory - Tetrahedral and Square Planar Complexes
Crystal field theory (CFT) is applicable to molecules in geometries other than octahedral. In octahedral complexes, the lobes of the dx2−y2 and dz2 orbitals point directly at the ligands. For tetrahedral complexes, the d orbitals remain in place, but with only four ligands located between the axes. None of the orbitals points directly at the tetrahedral ligands. However, the dx2−y2 and dz2 orbitals (along the Cartesian axes) overlap with the ligands less than the dxy,...
Chi-square Distribution
Residuals and Least-Squares Property
If the observed data point lies above the line, the residual is positive, and the line underestimates the actual data value for y. If the observed data point lies below the line, the residual is negative, and the line overestimates the actual data value for y.
The process of fitting the best-fit...