Assessment of calcification presence in hydrophilic acrylic phakic intraocular lenses
Zoha Mian1, Max Liu, Ethan Driffill
1From the John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
Purpose:
To evaluate explanted hydrophilic acrylic phakic intraocular lenses (pIOLs) for evidence of calcification, specifically the implantable collamer lens (ICL) and the I-CARE.
Setting:
Intermountain Ocular Research Center, John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.
Design:
Retrospective laboratory analysis of explanted pIOLs.
Methods:
Demographic data from the explanted cases were collected through a retrospective chart review. Explanted hydrophilic acrylic pIOLs were examined by gross and light microscopy to evaluate optical clarity, material integrity, and presence of surface or subsurface deposits. A subset of lenses underwent histochemical staining (1% alizarin red) and surface analysis to detect calcium deposits. Additional lenses were analyzed with implant cytology and visible-light transmission testing to assess biological deposition and UV-blocking capacity.
Results:
84 pIOLs were analyzed, including 65 ICLs (mean implantation time 8.78 ± 7.23 years) and 19 I-CARE lenses (mean implantation time 12.99 ± 5.34 years). Trace to mild pigmentary deposition was present in 36 ICLs (55.4%) and 10 I-CARE lenses (52.6%). No opacification or calcification was observed on gross or microscopic evaluation, including peripheral haptic areas. Histochemical staining and surface analysis confirmed absence of calcium deposits. Implant cytology revealed proteinaceous material without cellular components. Light transmission ranged from 93.7% to 97.4%, and all lenses demonstrated effective UV blocking.
Conclusions:
This large series of explanted hydrophilic acrylic pIOLs showed no evidence of calcification after long-term implantation. These findings support the material stability and biocompatibility of hydrophilic acrylic pIOLs in vivo.
Related Concept Videos
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Nursing Assessment
The nurse collects all aspects of the patient's health in the initial assessment, establishing priorities for ongoing focused assessments...
Assessment of the Rectum and Anus
Rectal Inspection
Begin by inspecting the perianal and anal areas for color, texture, rashes,...
Pulse Assessment Sites
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps: