Use of polystyrene spin-coated compact discs for microimmunoassaying
Jesús Tamarit-López1, Sergi Morais, Rosa Puchades
1Departamento de Química, Instituto de Química Molecular Aplicada, Universidad Politécnica de Valencia, Camino de Vera s/n, 46071 Valencia, Spain.
Abstract:
The analytical potential of polystyrene (PS) spin-coated modified compact discs (CDs) surface as platforms for the development of microarray immunoassays is presented. The surface maintained the optical characteristics of compact discs, obtaining a transparent and smooth film polymer of 70 nm thickness, the track being read (lambda 780 nm) without errors in a commercial CD reader/writer. The analytical capability of the methodology was demonstrated through an analysis of a neurotoxic compound (2560 spots per disc), reaching 0.08 microg L(-1) as limit of detection. These figures demonstrate the enormous potential of using PS spin-coated compact discs in combination with CD players as an easy-to-operate and portable device to develop lab-on-a-disc analytical applications.