[Accuracy limits in IOL calculation: current status]
1Universitäts-Augenklinik Mainz. pr.preussner@uni-mainz.de
Purpose:
Overview over the single independent error contributions of IOL power calculation and their currently achievable lower limits.
Method:
Analysis of the reasons of avoidable and unavoidable single errors which contribute to the overall error: measuring errors of axial length and corneal radii; errors due to neglecting of relevant influences such as pupil width, asphericity of cornea and IOL and IOL geometry; calculation errors from inadequate calculation methods; estimation errors of postoperative IOL position; IOL manufacturing errors. The said error contributions are to be compared with the reproducibility error of the refraction. All calculations use a numerical raytracing based on the geometric-optical IOL manufacturing data.
Results:
Axial eye length with an error of approximately 0.2 D is no longer the dominating error if the measurements are performed by interferometry, the same is true for corneal readii in normal eyes. The latter, however, causes the dominant error in eyes after corneal refractive surgery ( approximately 1.5 D) if measured only by keratometry. This error can be avoided if a topographic measurement is included into the raytracing, and in some cases also the measurement of posterior corneal surface has to be included. Currently the dominant unavoidable error contribution results from the uncertainty of postoperative IOL position ( approximately 0.35 D).
Conclusion:
Some errors of classical IOL formulae can be avoided by raytracing. But if the total error threshold shall be below the error of refraction, the prediction accuracy of postoperative IOL position must be improved.
More Related Videos
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Controlled-Current Coulometry: Overview
Limits with Oscillating Discontinuities
Types of Limits I
Types of Limits II