Two-modulator generalized ellipsometry: experiment and calibration
Abstract:
A two-modulator generalized ellipsometer is described that is capable of measuring all 16 elements of a sample Mueller matrix with four measurements made at different azimuthal orientations of the polarization state generator and polarization state detector. If the sample can be described with a Mueller-Jones matrix, only a single measurement is needed. Only two calibration steps are needed to determine the fundamental operating parameters of the instrument. A reflection measurement from silicon is presented as an example, which illustrates that the elements of the Mueller-Jones matrix can be measured to an accuracy of ~0.1-0.2%.
Related Concept Videos
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...