A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Inappropriate ICD shocks for inappropriate reasons.
Rakesh K Pai1, Moeen Abedin, David A Rawling
1Department of Internal Medicine, Cardiology Section, University of Utah Health Sciences Center, Salt Lake City, UT USA. rpai@renoheart.com
Inappropriate implantable cardioverter-defibrillator (ICD) shocks can be caused by electromagnetic interference (EMI) and lead terminal reversal. Identifying these unusual causes is crucial for correcting the ICD shock problem.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Implantable cardioverter-defibrillators (ICDs) are critical for managing life-threatening arrhythmias.
- Inappropriate ICD shocks, delivering unnecessary therapy, can stem from various causes.
- Accurate diagnosis of shock etiology is vital for patient management and device optimization.
Purpose of the Study:
- To highlight two uncommon etiologies of inappropriate ICD shocks.
- To emphasize the role of electromagnetic interference (EMI) in triggering shocks.
- To report on lead terminal reversal as a cause of device malfunction.
Main Methods:
- Case report presentation of two patients experiencing inappropriate ICD shocks.
- Analysis of device diagnostics to identify the underlying cause of shocks.
- Review of lead integrity and electromagnetic compatibility.
Main Results:
- Both patients experienced inappropriate ICD shocks attributed to external electromagnetic interference (EMI).
- A subsequent finding revealed reversal of the proximal and distal DF-1 lead terminals in both cases.
- Correction of lead terminal orientation resolved the inappropriate shocks.
Conclusions:
- Electromagnetic interference (EMI) can mimic cardiac signals, leading to inappropriate ICD shocks.
- Reversal of ICD lead terminals is an unusual but significant cause of device malfunction.
- Thorough investigation of device function and lead integrity is essential when managing inappropriate ICD shocks.
Related Concept Videos
Blood Pressure Imbalances and Circulatory Shock
Blood Pressure: Hypertension and Hypotension
Normal blood pressure is 120/80 mm Hg. Elevated blood pressure is 120-129/under 80 mm Hg. Hypertension, warranting treatment at 130/80 mm Hg, is often asymptomatic and can lead to severe cardiovascular events, aneurysms, peripheral arterial disease, chronic renal disease, or cardiac...
Cardiopulmonary Resuscitation III: AED Use
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
Antiarrhythmic Drugs: Class I Agents as Sodium Channel Blockers
Class 1A Antiarrhythmic Drugs: These drugs work by moderately blocking sodium channels,...