New clinical validation method for automated sphygmomanometer: a proposal by Japan ISO-WG for sphygmomanometer
Osamu Shirasaki1, Yosuke Asou, Yukio Takahashi
1Omron Healthcare Co. Ltd., Yamanouchi-yamanoshita-cho, Ukyo-ku, Kyoto, Japan. osamu_shirasaki@ohq.omron.co.jp
Introduction:
Owing to fast or stepwise cuff deflation, or measuring at places other than the upper arm, the clinical accuracy of most recent automated sphygmomanometers (auto-BPMs) cannot be validated by one-arm simultaneous comparison, which would be the only accurate validation method based on auscultation. Two main alternative methods are provided by current standards, that is, two-arm simultaneous comparison (method 1) and one-arm sequential comparison (method 2); however, the accuracy of these validation methods might not be sufficient to compensate for the suspicious accuracy in lateral blood pressure (BP) differences (LD) and/or BP variations (BPV) between the device and reference readings. Thus, the Japan ISO-WG for sphygmomanometer standards has been studying a new method that might improve validation accuracy (method 3). The purpose of this study is to determine the appropriateness of method 3 by comparing immunity to LD and BPV with those of the current validation methods (methods 1 and 2).
Method:
The validation accuracy of the above three methods was assessed in human participants [N=120, 45+/-15.3 years (mean+/-SD)]. An oscillometric automated monitor, Omron HEM-762, was used as the tested device.
Results:
When compared with the others, methods 1 and 3 showed a smaller intra-individual standard deviation of device error (SD1), suggesting their higher reproducibility of validation. The SD1 by method 2 (P=0.004) significantly correlated with the participant's BP, supporting our hypothesis that the increased SD of device error by method 2 is at least partially caused by essential BPV. Method 3 showed a significantly (P=0.0044) smaller interparticipant SD of device error (SD2), suggesting its higher interparticipant consistency of validation.
Conclusion:
Among the methods of validation of the clinical accuracy of auto-BPMs, method 3, which showed the highest reproducibility and highest interparticipant consistency, can be proposed as being the most appropriate.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.