A novel hemoglobin, Bonn, causes falsely decreased oxygen saturation measurements in pulse oximetry
Berndt Zur1, Andreas Hornung, Johannes Breuer
1Department of Clinical Biochemistry and Division of Pediatric Cardiology, University of Bonn, Bonn, Germany. berndt.zur@ukb.uni-bonn.de
Background:
A 4-year-old boy and his father exhibited low oxygen saturation measured transcutaneously by pulse oximetry, a finding that could not be confirmed by arterial blood gas analysis. Both patients exhibited slight hemolysis in their blood, and the boy had a microcytic anemia. There was no evidence of hypoxemia or methemoglobinemia. Despite the normal results from the arterial blood gas analysis, a right-to-left-shunt was assumed in the boy until a cardiology examination excluded this diagnosis. Sleep apnea syndrome was suspected in the father and treated with nocturnal positive pressure respiration based on the low oxygen saturation values obtained with pulse oximetry. Only after consultation with our laboratory was a hemoglobin variant suspected and investigated.
Methods:
We performed hemoglobin protein analysis by HPLC, electrophoretic separation, and spectrophotometry and DNA sequence analysis of the alpha-globin gene.
Results:
Both HPLC chromatographic separation and alkaline electrophoresis revealed a unique hemoglobin peak. In both patients, alpha-globin gene sequencing revealed a mutation resulting in a histidine-to-aspartatic acid substitution at position alpha87. The low oxygen saturation measurement by pulse oximetry was due to hemoglobin Bonn oxyhemoglobin having an absorption peak at 668 nm, near the 660 nm measured by pulse oximeters.
Conclusion:
Hemoglobin Bonn is a novel hemoglobin variant of the proximal alpha-globin that results in falsely low oxygen saturation measurements with pulse oximetry.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Transport in the Blood
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Hypoxia
Types of Hypoxia
There are four primary types of hypoxia, each resulting from a different cause:
1. Anemic hypoxia: This type occurs due to insufficient oxygen delivery caused by a lack of red blood cells (RBCs) or RBCs with abnormal or...
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...