Experimental comparison of time-of-flight mass analysis with magnetic mass analysis
V I Gushenets1, A S Bugaev, E M Oks
1High Current Electronics Institute, Russian Academy of Sciences, Tomsk, Russia.
Abstract:
A series of experiments was carried out in which both a magnetic analyzer (mass separator) and a time-of-flight (TOF) spectrometer were used for ion charge/mass spectral analysis of the ion beam formed by a dc Bernas ion source made for semiconductor implantation. The TOF analyzer was a detachable device that provides rapid analysis of charge-to-mass composition of moderate energy ion beams. The magnetic analyzer was a massive device using a 90 degrees -sector bending magnet with radius of the central orbit of 35 cm. Comparison of these two methods for measuring ion beam composition shows good agreement.