Errors involved in the existing B-term expressions for the longitudinal diffusion in fully porous chromatographic
K Broeckhoven1, D Cabooter, F Lynen
1Vrije Universiteit Brussel, Department of Chemical Engineering, Pleinlaan 2, 1050 Brussels, Belgium.
Abstract:
Peak parking experiments have been performed on three RP-HPLC different columns, using two different components and a variable mobile phase composition. The aim of the study was to investigate whether the B-term diffusion expressions currently used in the literature (which are all Knox-type models) should be replaced by the effective diffusion expressions that have been developed in the frame of the effective medium theory (EMT). Although the EMT-expressions are not fully accurate either (the mathematics of the complex interactions between different diffusion zones that are in close contact are too demanding to catch them in an exact analytical expression), they at least are physically sound and do not violate Maxwell's basic law of diffusion. Further they also provide a much better approximation of the numerically calculated effective diffusivity in the theoretical test situation considered in part I. The present study shows that the values of the surface or stationary phase diffusion coefficient that are derived from peak parking models can depend heavily on the employed B-term model. The EMT-based B-term expressions lead to values of the surface diffusion coefficient that vary much less strongly with the phase retention factor than if one of the Knox-type models is used to analyze the data. This implies that, since all peak parking experiments that have been performed in the past have all been interpreted with a Knox-type model, the conclusions that have been drawn from these studies should all be moderated or at least revisited.
Related Concept Videos
Diffusion on Chromatography Columns
Longitudinal diffusion occurs when the solute molecules in the mobile phase diffuse from the more concentrated center of the chromatographic band to the more dilute regions on either side, both towards and against the flow direction. This...
Column Efficiency: Rate Theory
During elution, a solute molecule experiences numerous transitions between stationary and mobile phases, exhibiting irregular residence times in...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Ion-Exchange Chromatography
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
Chromatography: Introduction
The phase in which the compounds linger or on which the compounds adsorb is called the stationary phase, whereas the mobile phase is the solvent that carries the solutes to be analyzed. In traditional column chromatography, the mixture flows through the stationary phase, and the compounds partition between the stationary and mobile phases...