Considerations for the use of ultra-high pressures in liquid chromatography for 2.1mm inner diameter columns
1Vrije Universiteit Brussel, Department of Chemical Engineering, Pleinlaan 2, 1050 Brussels, Belgium.
Abstract:
The current contribution investigates the effects of viscous heat dissipation in chromatographic columns (with an emphasis on so-called narrow bore columns with an inner diameter of 2.1mm) using numerical simulations of the temperature and velocity profiles and the resulting band broadening, for the first time at operating pressures up to 2000bar. When operating columns under well-thermostatted conditions to maintain a constant temperature of the mobile phase, a dramatic increase in plate heights can be observed that voids any advantage one could expect from the possibility to use smaller particles offered by the increased pressure limit. It is also clearly demonstrated that, even when the column is not temperature controlled, the backflow of heat along the wall can causes a significant loss in performance under standard operating conditions in a still air oven. It is found that for operating pressure above 1250bar, a significant (relative to the typical column performance) contribution to the observed plate height will be caused by viscous heating effects, which increases with increasing temperature dependency of the retention factor. In addition, unprecedented experimental measurements of the temperature effects at an operating pressure up to 2600bar were performed on a 10cm long, 2.1mm ID column showing a dramatic temperature increase up to 60°C relative to the inlet temperature when using methanol as a mobile phase.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Principles Of Column Chromatography
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
High-Performance Liquid Chromatography: Instrumentation