A Methodological Approach to Non-invasive Assessments of Vascular Function and Morphology
Published on: February 7, 2015
In vivo vibroacoustography of large peripheral arteries
Cristina Pislaru1, Birgit Kantor, Randall R Kinnick
1Department of Physiology and Biomedical Engineering, Mayo Clinic College of Medicine, Rochester, Minnesota 55905, USA. pislaru.cristina@mayo.edu
Objective:
Vibroacoustography allows imaging of objects on the basis of their acoustic signal emitted during low-frequency (kHz) vibrations produced by 2 intersecting ultrasound beams at slightly different frequencies. This study tested the feasibility of using vibroacoustography to distinguish between normal and calcified femoral arteries in a pig model.
Materials And Methods:
Thirteen normal porcine femoral arteries, 7 with experimentally induced arterial calcifications, and 1 control artery injected with saline only were scanned in vivo. Images were obtained at 45 kHz using a 3 MHz confocal transducer. The acoustic emission signal was detected with a hydrophone placed on the animal's limb. Images were reconstructed on the basis of the amplitude of the acoustic emission signal. Vessel patency, vessel dimensions, and the extent of calcified plaques were confirmed in vivo by angiography and conventional ultrasound. Excised arteries were reexamined with vibroacoustography, X-ray radiography, and histology.
Results:
In vivo, vibroacoustography produced high-resolution, speckle-free images with a high level of anatomic detail. Measurements of femoral artery diameter were similar by vibroacoustography and conventional ultrasound (mean difference +/- SD, 0.1 +/- 0.4 mm). Calcified plaque area measured by different methods was comparable (vibroacoustography, in vivo: 1.0 +/- 0.9 cm; vibroacoustography in vitro: 1.1 +/- 0.6 cm2; X-ray radiography: 0.9 +/- 0.6 cm2). The reproducibility of measurements was high. Sensitivity and specificity for detecting calcifications were 100% and 86%, respectively, and positive and negative predictive values were 77% and 100%, respectively.
Conclusions:
Vibroacoustography provides accurate and reproducible measurements of femoral arteries and vascular calcifications in living animals.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Imaging Studies II: Ultrasonography
Imaging Studies for Cardiovascular System IV: CMRI
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Imaging Studies VII: Vascular Imaging