Related Experiment Video
Updated: Jul 6, 2026

Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
Time-resolved three-dimensional imaging of the left atrium and pulmonary veins in the interventional suite--a
Amin Al-Ahmad1, Lars Wigström, Dominique Sandner-Porkristl
1Cardiac Arrhythmia Service, Stanford University Medical School, Stanford, California, USA. aalahmad@cvmed.stanford.edu
Background:
Cardiac computed tomography (CT) is commonly used to visualize left atrial (LA) anatomy for ablation of atrial fibrillation. We have developed a new imaging technique that allows acquisition and visualization of three-dimensional (3D) cardiac images in the catheter lab.
Objective:
We sought to compare LA and pulmonary vein (PV) dimensions acquired using gated multisweep rotational fluoroscopy (C-arm CT) system and multislice computed tomography (MSCT) in an in vivo porcine model.
Methods:
A Siemens AXIOM Artis dTA C-arm system (Siemens AG, Medical Solutions) was modified to allow acquisition of four bidirectional sweeps during synchronized acquisition of the electrocardiogram signal to allow retrospective gating. C-arm CT image volumes were then reconstructed. Gated MSCT (SOMATOM Sensation 16 and 64, Siemens AG, Medical Solutions) and C-arm CT images were acquired in six animals. The two main PV diameters were measured in orthogonal axes. LA volumes were calculated. C-arm CT measurements were compared with the MSCT measurements.
Results:
The average PV diameters using the C-arm CT were 2.24 x 1.35 cm, versus 2.27 x 1.38 cm for CT. The average difference was 0.034 cm (1.9%) between the C-arm CT and standard CT. The average LA volume using MSCT was 49.1 +/- 12.7 cm(3), as compared with 51.0 +/- 8.7 cm(3) obtained by the C-arm CT. The average difference between the C-arm CT and the MSCT was 1.9 cm(3) (3.7%). There were no significant differences in either the PV or LA measurements.
Conclusions:
Visualization of 3D cardiac anatomy during ablation procedures is possible and highly accurate. The 3D cardiac reconstructions acquired during ablation procedures will be valuable for procedural planning and guidance.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization III: Left Heart Catheterization
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...