Related Experiment Video
Updated: Jul 6, 2026

Fabrication and Characterization of a Conformal Skin-like Electronic System for Quantitative, Cutaneous Wound Management
Published on: September 2, 2015
Carbon fibre composites: integrated electrochemical sensors for wound management
Duncan Sharp1, Stephen Forsythe, James Davis
1School of Science and Technology, Nottingham Trent University, Nottingham, NG11 8NS, UK. duncan.sharp@ntu.ac.uk
Abstract:
The applicability of employing a carbon fibre mesh as an electrochemical sensing substructure for assessing urate transformations within wound exudates is evaluated. Prototype sensor assemblies have been designed and their response characteristics towards uric acid and other common physiological components are detailed. Modification of the carbon fibre sensor through surface anodization and the application of cellulose acetate permselective barriers have been shown to lead to optimized responses and much greater sensitivity (1440% increase) and specificity. These could enable the accurate periodic monitoring of uric acid in wound fluid. The performance characteristics of the composite sensors in whole blood, serum and blister fluid have been investigated.