Objective:
To test the efficiency of the Bird Mark 7 ventilator adapted to deliver continuous positive airway pressure (CPAP) in noninvasive positive pressure ventilation.
Methods:
This was an experimental study using a mechanical model of the respiratory system. A Bird Mark 7 ventilator was supplied with 400 and 500 kPa and tested at CPAP of 5, 10 and 15 cmH2O. The following variables were analyzed: difference between the preset CPAP and the CPAP actually attained CPAP (trueCPAP); area of airway pressure at the CPAP level employed (AREA CPAP); and tidal volume generated.
Results:
Adapting the Bird Mark 7 to offer CPAP achieved the expected tidal volume in all situations of inspiratory effort (normal or high), ventilator pressure supply (400 or 500 kPa) and CPAP value (5, 10 or 15 cmH2O). At a CPAP of 5 or 10 cmH2O, the trueCPAP was near the preset level, and the AREA CPAP was near zero. However, at a CPAP of 15 cmH2O, the value remained below the preset, and the AREA CPAP was high.
Conclusion:
The efficiency of Bird Mark 7 adaptation in offering CPAP was satisfactory at 5 and 10 cmH2O but insufficient at 15 cmH2O. If adapted as described in our study, the Bird Mark 7 might be an option for offering CPAP up to 10 cmH2O in areas where little or no equipment is available.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Cardiopulmonary Resuscitation II: ACLS Airway Management
Suctioning the Nasopharyngeal Airway
Equipment Required
Sleep Apnea
The condition is more prevalent among...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...