Portable ultrasonography and bladder volume accuracy--a comparative study using three-dimensional ultrasonography.
Khurshid R Ghani1, James Pilcher, David Rowland
1Department of Urology, St. George's Hospital Medical School, London, United Kingdom. krghani@gmail.com
The three-dimensional ultrasound (3D-US) system demonstrated the highest accuracy and clinical agreement for bladder volume measurement compared to portable bladder scanners. The Bardscan underestimated voided volumes, while Bladderscan and 3D-US showed no significant differences.
Area of Science:
- Medical Imaging
- Diagnostic Ultrasound
- Urology
Background:
- Accurate bladder volume measurement is crucial for clinical decision-making.
- Portable bladder scanners offer convenience but vary in accuracy.
- Three-dimensional ultrasound (3D-US) presents an alternative for volumetric assessment.
Purpose of the Study:
- To compare the accuracy and agreement of bladder volume measurements between two portable bladder scanners (Bladderscan and Bardscan) and a 3D-US system.
- To evaluate the reliability of these devices in healthy volunteers.
Main Methods:
- Fifty healthy volunteers underwent sequential scanning with Bladderscan BVI 3000, Bardscan, and 3D-US (HDI 4000).
- Ultrasound-derived bladder volumes were compared against actual voided volumes.
- Subjects with measurable post-void residual volumes were excluded from analysis.
Main Results:
- All three methods showed significant correlation (r=0.97-0.99) between voided and ultrasound volumes.
- Bladderscan and 3D-US showed no significant difference from voided volumes.
- Bardscan significantly underestimated voided volume by a mean of 21.4 mL.
- Bland-Altman limits of agreement were narrowest for 3D-US (-28.9 to 40.0 mL).
Conclusions:
- The three-dimensional ultrasound (3D-US) system offers the greatest accuracy and clinical agreement for bladder volume calculation.
- While portable, the Bardscan is less accurate than Bladderscan and underestimates bladder volume.
- Bladderscan and 3D-US provide reliable bladder volume estimations.
Related Concept Videos
Imaging Studies II: Ultrasonography
Imaging Studies VI: Voiding Cystourethrography and Cystography
Urodynamic Studies: Uroflowmetry
Imaging Studies III: Computed Tomography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Ultrasonography
During an ultrasonography procedure, a handheld device called a...