Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Imaging Studies for Cardiovascular System III: X-Ray01:20

Imaging Studies for Cardiovascular System III: X-Ray

The most common cardiovascular diagnostic test is an X-ray. It produces images of the heart, blood vessels, and adjacent structures.
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Mitral Valve Leaflet Modification: A Working Group Position Statement on Best Practices and Step-by-Step Guide.

JACC. Cardiovascular interventions·2026
Same author

UNICORN During Transcatheter Aortic Valve Replacement in Native and Bioprosthetic Aortic Valves: A Multicenter Study.

JACC. Cardiovascular interventions·2026
Same author

Failure Mechanisms in Off-Label Electrosurgery Devices for BASILICA, LAMPOON, and SESAME.

JACC. Cardiovascular interventions·2026
Same author

Aortic Valve Leaflet Modification: A Working Group Position Statement on Best Practices and Step-by-Step Guide.

JACC. Cardiovascular interventions·2026
Same author

Current Evidence, Emerging Evidence, and Decision-Making: An Algorithm for Leaflet Modification Procedures.

JACC. Cardiovascular interventions·2026
Same author

Adapting Interventional Echocardiography to New Techniques: Echocardiographic Guidance for Transcatheter Myotomy Using Septal Scoring Along the Midline Endocardium.

Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography·2026

Related Experiment Video

Updated: Jul 6, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

A practical global distortion correction method for an image intensifier based x-ray fluoroscopy system.

Luis F Gutiérrez1, Cengizhan Ozturk, Elliot R McVeigh

  • 1Department of Biomedical Engineering, Johns Hopkins University, Baltimore, Maryland 21205, USA. luis.gutierrez@philips.com

Medical Physics
|April 15, 2008
PubMed
Summary
This summary is machine-generated.

A new method corrects X-ray image distortion from image intensifiers (II) on C-arms. This technique improves accuracy across various C-arm orientations, crucial for medical imaging guidance and planning.

More Related Videos

Measuring the Complete-arch Distortion of an Optical Dental Impression
06:51

Measuring the Complete-arch Distortion of an Optical Dental Impression

Published on: May 30, 2019

Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration
07:03

Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration

Published on: February 23, 2017

Related Experiment Videos

Last Updated: Jul 6, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

Measuring the Complete-arch Distortion of an Optical Dental Impression
06:51

Measuring the Complete-arch Distortion of an Optical Dental Impression

Published on: May 30, 2019

Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration
07:03

Medical-grade Sterilizable Target for Fluid-immersed Fetoscope Optical Distortion Calibration

Published on: February 23, 2017

Area of Science:

  • Medical Imaging
  • Image Processing
  • Radiological Technology

Background:

  • X-ray images from image intensifiers (II) have distortions due to internal and external factors.
  • Distortion varies with the II's orientation, especially on C-arms with multiple degrees of freedom.
  • Existing distortion correction methods often require extensive C-arm orientation sampling.

Purpose of the Study:

  • To develop a novel method for correcting image distortion in X-ray systems with image intensifiers.
  • To enable accurate distortion correction across a wide range of C-arm movements.
  • To enhance image quality for image-guided interventions and treatment planning.

Main Methods:

  • A new method smooths segmented grid phantom vertices trajectories before solving the 2D warping problem.
  • An alternative approach involves global distortion correction followed by smoothing polynomial coefficients as functions of C-arm orientation parameters.
  • The method was extended to three degrees of freedom: primary angle (alpha), secondary angle (beta), and source-to-intensifier distance (lambda).

Main Results:

  • The new method achieved a root-mean-square (RMS) residual error of 0.22 pixels for arbitrary alpha.
  • With three degrees of freedom (alpha, beta, lambda), the RMS residual error was 0.33 pixels using only 75 calibration images.
  • Characterization covered alpha: +/- 45 degrees, beta: +/- 36 degrees, and lambda: 98-118 cm.

Conclusions:

  • The developed method effectively corrects image intensifier distortion over a broad range of C-arm orientations.
  • This technique offers improved accuracy for applications requiring precise image guidance, such as catheter-based interventions and brachytherapy.
  • The method's efficiency in calibration (75 images) makes it practical for clinical use.