Wavelet analysis of valve closing sound detects malfunction of bileaflet mechanical valve
Hiroshi Sugiki1, Toshifumi Murashita, Norihiko Shiiya
1Department of Cardiovascular Surgery, Hokkaido University Postgraduate School of Medicine, North-14, West-4, Sapporo, 060-8648, Japan. Hsugiki@aol.com
Abstract:
Several studies have reported the asynchronous closure of normal bileaflet valves (NBVs), resulting in a split in its closing sound; however, the clinical significance of this split has never been studied in malfunctioning bileaflet valves (MBVs). The study comprised 218 valves in 184 patients, including normal monoleaflet valves (n = 10), NBVs (n = 198), and MBVs (n = 10). Valve function was confirmed by cinefluoroscopy prior to analysis of the valve sound by the Morlet continuous wavelet transform (CWT). The split interval (SI) for each heartbeat was measured, and the coefficient of variation (CV) of its mean (valve SI) was calculated as a parameter for the fluctuation of the SI. The CWT of monoleaflet valves showed a single spike, whereas NBVs exhibited a clear split. There was no significant difference in valve SI between the aortic and mitral positions; however, the mean of the CV was significantly greater in the mitral position (n = 90, 0.507 +/- 0.254) than in aortic position (n = 108, 0.353 +/- 0.228, P = 0.000045). The split was not found in six (aortic; three, mitral; three) of ten patients with MBVs. The other four patients had a distinct split, but the CV was significantly lower for MBVs (0.138 +/- 0.105) than for NBVs (0.343 +/- 0.221, P = 0.042). Receiver-operating characteristics analysis demonstrated the cutoff line of the CV to be 0.112 for detecting malfunctioning aortic valves with the highest accuracy of 86.1%. This new system using the Morlet CWT can detect MBVs. It will be a useful modality for screening the function of bileaflet valves.
More Related Videos
06:45Design and Application of a Fault Detection Method Based on Adaptive Filters and Rotational Speed Estimation for an Electro-Hydrostatic Actuator
Published on: October 28, 2022
06:56An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
Related Concept Videos
Heart Sounds
Auscultation is the process of listening to these internal body sounds using a stethoscope. The heart produces four types of sounds, but only two—S1 and S2—can usually be heard with a stethoscope.
S1, also known as the "lub" sound, is caused by the closure of atrioventricular (A-V) valves at the...
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops:
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.