Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Characteristics of mechanical heart valve cavitation in a pneumatic ventricular assist device
Hwansung Lee1, Yoshiyuki Taenaka
1Department of Artificial Organs, Research Institute, National Cardiovascular Center, Osaka, Japan. hslee@ri.ncvc.go.jp
Abstract:
In previous studies, we investigated the mechanism of mechanical heart valve (MHV) cavitation and cavitation intensity with a nonsynchronized experiment system. Our group is currently developing a pneumatic ventricular assist device (PVAD), and in this study we investigated MHV cavitation intensity in the PVAD using a synchronized analysis of the cavitation images and the acoustic signal of cavitation bubbles. A 23-mm Medtronic Hall valve with an opening angle of 70 degrees was mounted in the mitral position of the PVAD after removing the sewing ring. A function generator provided a square signal, which used the trigger signal of the electrocardiogram R wave (ECG-R) mode of the control-drive console for circulatory support. This square signal was delayed by a delay circuit and was then used as the trigger signal for a pressure sensor and a high-speed video camera. The data were stored using a digital oscilloscope at a 1-MHz sampling rate, and then the pressure signal was band-pass filtered between 35 and 200 kHz using a digital filter. The band-pass filtered root mean squared (RMS) pressure and cavitation cycle duration were used as an index of cavitation intensity. The cavitation bubbles were concentrated at the valve stop, and the cavitation cycle duration and RMS pressure increased as the heart rate and driving pressure increased. At the low valve-closing velocity, bubble cavitation was observed near the valve stop. However, at the fast valve-closing velocity, cloud cavitation was observed. A high-frequency signal wave was generated when the bubbles collapsed. The cavitation cycle duration and RMS pressure increased as the valve-closing velocity increased linearly.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Aortic Regurgitation II: Clinical Features and Diagnostic Tests