Quantification of coronary plaque by 64-slice computed tomography: a comparison with quantitative intracoronary
Masato Otsuka1, Nico Bruining, Niels C Van Pelt
1Department of Cardiology, Erasmus Medical Center, Rotterdam, The Netherlands.
Background:
Noninvasive assessment of coronary atherosclerotic plaque may be useful for risk stratification and treatment of atherosclerosis.
Materials And Methods:
We studied 47 patients to investigate the accuracy of coronary plaque volume measurement acquired with 64-slice multislice computed tomography (MSCT), using newly developed quantification software, when compared with quantitative intracoronary ultrasound (QCU). Quantitative MSCT coronary angiography (QMSCT-CA) was performed to determine plaque volume for a matched region of interest (regional plaque burden) and in significant plaque defined as a plaque with > or =50% area obstruction in QCU, and compared with QCU. Dataset with image blurring and heavy calcification were excluded from analysis.
Results:
In 100 comparable regions of interest, regional plaque burden was highly correlated (coefficient r = 0.96; P < 0.001) between QCU and QMSCT-CA, but QMSCT-CA overestimated the plaque burden by a mean difference of 7 +/- 33 mm3 (P = 0.03). In 76 significant plaques detected within the regions of interest, plaque volume determined by QMSCT-CA was highly correlated (r = 0.98; P < 0.001) with a slight underestimation of 2 +/- 17 mm3 (P = not significant) when compared with QCU. Calcified and mixed plaque volume was slightly overestimated by 4 +/- 19 mm3 (P = ns) and noncalcified plaque volume was significantly underestimated by 9 +/- 11 mm3 (P < 0.001) with QMSCT-CA. Overall, the limits of agreement for plaque burden/volume measurement between QCU and QMSCT-CA were relatively large. Reproducibility for the measurements of regional plaque burden with QMSCT-CA was good, with a mean intraobserver and interobserver variability of 0% +/- 16% and 4% +/- 24%, respectively.
Conclusions:
Quantification of coronary plaque within selected proximal or middle coronary segments without image blurring and heavy calcification with 64-slice CT was moderately accurate with respect to intravascular ultrasound and demonstrated good reproducibility. Further improvement in CT resolution is required for more reliable measurement of coronary plaques using quantification software.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...