Estimation of the aortic pressure waveform and beat-to-beat relative cardiac output changes from multiple peripheral
Gokul Swamy1, Ramakrishna Mukkamala
1Department of Electrical and Computer Engineering, Michigan State University, East Lansing, MI 48824, USA. swamygok@msu.edu
Abstract:
We introduce a patient- and time-specific technique to estimate the clinically more relevant aortic pressure (AP) waveform and beat-to-beat relative changes in cardiac output (CO) from multiple peripheral artery pressure (PAP) waveforms distorted by wave reflections. The basic idea of the technique is to first estimate the AP waveform by applying a new multichannel blind system identification method that we have developed (rather than the conventional generalized transfer function) to the PAP waveforms and then estimate the beat-to-beat proportional CO by fitting a Windkessel model to the estimated waveform in which wave distortion should be attenuated. We present an evaluation of the technique with respect to four swine datasets including simultaneous measurements of two peripheral AP waveforms, a reference AP waveform, and reference aortic flow probe CO during diverse hemodynamic interventions. Our results show an overall AP waveform error of 3.5 mmHg and an overall beat-to-beat CO error of 12.9% (after a single CO calibration in each animal). These estimation errors represent substantial improvements compared to those obtained with several alternative PAP waveform analysis techniques. With further successful testing, the new technique may ultimately be employed for automated and less invasive monitoring of central hemodynamics in various cardiovascular patients.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)