Related Experiment Video
Updated: Jul 5, 2026

An Image Guided Transapical Mitral Valve Leaflet Puncture Model of Controlled Volume Overload from Mitral Regurgitation in the Rat
Published on: May 19, 2020
Acute reduction of functional mitral regurgitation in canine model using an epicardial device
Masatoshi Akiyama1, Zoran B Popović, Keiji Kamohara
1Department of Biomedical Engineering, Lerner Research Institute, Cleveland, Ohio, USA.
Purpose:
This study evaluated the short-term feasibility of a novel epicardial device that treats functional mitral regurgitation by simultaneously changing the mitral and the left ventricular geometry.
Description:
We implanted a prototype device that consists of 2 tissue anchors, a deflector, and a flexible tightening chord in 7 mongrel dogs with heart failure and functional mitral regurgitation induced by rapid ventricular pacing. Hemodynamic and echocardiographic data were obtained before and after device implantation.
Evaluation:
The device acutely reduced the mitral regurgitation grade from 3.2 +/- 0.3 to 0.9 +/- 0.5 (p < 0.001). Left ventricular end-diastolic volume (79.6 +/- 23.6 to 61.2 +/- 16.9 mL; p = 0.004) and end-systolic volume (63.1 +/- 17.3 to 49.2 +/- 12.3 mL; p = 0.006) decreased substantially. End-systolic elastance significantly increased from 1.9 +/- 1.0 to 2.6 +/- 1.4 mm Hg/mL (p = 0.02). Device implantation did not alter coronary perfusion.
Conclusions:
The epicardial device acutely reduced functional mitral regurgitation and improved left ventricular geometry. Further studies are required to demonstrate the long-term safety and efficacy of this concept.
More Related Videos
Related Concept Videos
Mitral Regurgitation III: Medical Management
Mitral Stenosis III: Medical Management