Autoclaving and battery capacity influence on laser fluorescence measurements
Renata Maciel Rocha-Cabral1, Fausto Medeiros Mendes, Fabiana Miura
1Centro de Lasers e Aplicações, Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, São Paulo, Brazil.
Aim:
To evaluate the influence of probe tip autoclaving and depleted alkaline batteries on laser fluorescence (LF-DIAGNOdent) device performance.
Material And Methods:
One-hundred-and-twenty occlusal sites were analyzed with an LF device in nine conditions: No autoclaved probe and probe after 5, 10, 15, 20, 25, 30, 40, and 50 autoclaving cycles. Subsequently, the performance of the device was analyzed with: New batteries, batteries with 1.49/1.39 V, 1.38/1.37 V, 1.36/1.34 V, 1.33/1.32 V, and lower than 1.32 V. LF values, sensitivity, specificity, accuracy, and area under the ROC curve were compared.
Results:
In the probe study, sensitivity was lower after 50 sterilizing cycles, though specificity was higher than the assessment performed using a new tip. In the batteries study, specificity was higher for depleted batteries, but LF performance did not differ significantly among the groups.
Conclusion:
Batteries do not significantly influence device performance, but consecutive sterilization of probes in autoclave alters readings, downgrading its performance.