Related Experiment Video
Updated: Jul 5, 2026

Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
A novel approach to treatment of unexpected vertebroplasty complication
Jonathan M Hard1, Roger L Gonda, Saumil R Kadakia
1Department of Radiology, Providence Hospital, 16001 West Nine Mile Road, Southfield, MI 48075, USA. jonathan_hard2001@yahoo.com
Abstract:
We report a novel approach to treatment of an unexpected complication of vertebroplasty. During initial positioning of transpedicular vertebroplasty needles in a 73-year-old woman, we encountered unexpected ease of needle advancement resulting in progression of the needle through the anterior cortex of the thoracic vertebral body. The transpedicular needle advanced into the mediastinum and, presumably, the adventitia of the descending thoracic aorta. Administration of polymethylmethacrylate cement was performed to tamponade bleeding at the time of the procedure and to reduce the risk of potential delayed bleeding complications within the mediastinum. The treatment was successful and the patient had no immediate or delayed complications as a result of the unintended needle advancement.