Field testing of lake water chemistry with a portable and an AUV-based mass spectrometer
Harry F Hemond1, Amy V Mueller, Michael Hemond
1Department of Civil and Environmental Engineering, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA. hfhemond@mit.edu
Abstract:
Two mass spectrometers (MS) are tested for the measurement of volatile substances, such as hydrocarbons and metabolic gases, in natural waters. KOALA is a backpackable MS operated from above the water surface, in which samples are pumped through a flow cell using a syringe. NEREUS is an underwater instrument hosted by an autonomous underwater vehicle (AUV) that is linked to a communications network to provide chemical data in real time. The mass analyzers of the two MS are nearly identical cycloids, and both use flat-plate membrane inlets. Testing took place in an eutrophic, thermally stratified lake exhibiting steep chemical gradients and significant levels of methane. KOALA provided rapid multispecies analysis of dissolved gases, with a detection limit for methane of 0.1 ppm (readily extendable to 0.01 ppm) and savings of time of at least a factor of 10 compared to that of conventional analysis. The AUV-mounted NEREUS additionally provided rapid spatial coverage and the capability of performing chemical surveys autonomously. Tests demonstrated the need for temperature control of a membrane inlet when steep thermal gradients are present in a water body, as well as the benefits of co-locating all sensors on the AUV to avoid interference from chemically different waters entering and draining from the free-flooding outer hull. The ability to measure dissolved volatiles provided by MS offers potential for complementarity with ionic sensors in the study of natural waters, such as in the case of the carbonate system.
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Testing Water Quality
High-Performance Liquid Chromatography: Types of Detectors
Atomic Emission Spectroscopy: Lab
Gravimetry: Overview
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.