Towards an automated, standardized protocol for determination of equilibrium potential of ion-selective electrodes
Amy V Mueller1, Harold F Hemond
1Massachusetts Institute of Technology, 15 Vassar St., Cambridge, MA, 02139, USA. amym@mit.edu
Abstract:
An automated real-time method for determination of ISE steady state value and response time is developed, following most recent IUPAC recommendations. Specifically, detection of the 'steady state' is related to (1) the time derivative of the emf as it reaches a limiting value (ΔE/Δt(limit), e.g., 0.1-1.0 mV min(-1)) and (2) the duration of time for which the absolute value of the time derivative remains less than this limiting value (stability window, denoted win(st)). A suite of representative ISEs, including glass, solid state, and polymer-based electrodes, is examined to determine sensitivity of results to parameterization choice. Measurements taken over a wide range of concentration values and in un-processed samples (i.e., without use of ionic strength adjustment) provide insight into behavior of ISEs in applications where analyte concentrations span a wide range and/or sample pre-processing may not be an option, e.g., use of sensors for in situ environmental sampling. Results show that declared steady state emf is strongly sensitive to variations in ΔE/Δt(limit) but relatively unaffected by changes in the stability window when win(st) ≥30 s. Linearity of calibration curves produced, quantified by root mean squared error (RMSE) against a linear fit, improves as ΔE/Δt(limit) decreases, however the percentage of measurements which reach a declared steady state within the prescribed sample window (∼6.5 min) falls with corresponding decreases in the ΔE/Δt(limit) parameter. Response time, defined as the time required to reach declared steady emf, is also a strong function of parameterization. Dependence of response times on sample composition and/or ISE membrane composition and type are also discussed; results for ISEs in samples comprised exclusively of interfering ions are included. In general, limiting emf derivatives of {0.25-0.4 mV min(-1)} and stability windows of {30-40s} achieve both good analytical accuracy and compliance with potentially short sampling window requirements. Methodology based on use of these parameters can improve sampling speed and accuracy as well as promote inter-comparison of data and ISE characterizations among research teams.
Related Concept Videos
Potentiometry: Membrane Electrodes
Potentiometry: Overview
Standard Electrode Potentials
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...
Ion-Exchange Chromatography