Evaluation of exposure dose to patients undergoing catheter ablation procedures--a phantom study
S Seguchi1, T Aoyama, S Koyama
1Graduate School of Medicine, Nagoya University, Higashi-ku, Nagoya, Japan. seguchi@nagoya2.jrc.or.jp
Abstract:
The aim of this study was to evaluate entrance skin dose (ESD), organ dose and effective dose to patients undergoing catheter ablation for cardiac arrhythmias, based on the dosimetry in an anthropomorphic phantom. ESD values associated with mean fluoroscopy time and digital cine frames were in a range of 0.12-0.30 Gy in right anterior oblique (RAO) and 0.05-0.40 Gy in left anterior oblique (LAO) projection, the values which were less than a threshold dose of 2 Gy for the onset of skin injury. Organs that received high doses in ablation procedures were lung, followed by bone surface, esophagus, liver and red bone marrow. Doses for lung were 24.8-122.7 mGy, and effective doses were 7.9-34.8 mSv for mean fluoroscopy time of 23.4-92.3 min and digital cine frames of 263-511. Conversion coefficients of dose-area product (DAP) to ESD were 8.7 mGy/(Gy cm(2)) in RAO and 7.4 mGy/(Gy cm(2)) in LAO projection. The coefficients of DAP to the effective dose were 0.37 mSv/(Gy cm(2)) in RAO, and 0.41 mSv/(Gy cm(2)) in LAO projection. These coefficients enabled us to estimate patient exposure in real time by using monitored values of DAP.
More Related Videos
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System V: CT