Comparison of arterial closure devices in antegrade and retrograde punctures

Albert Ho Yuen Chiu1, Ross Vander Wal, Kenneth Tee

  • 1Department of Radiology, Sir Charles Gairdner Hospital, Perth, Australia. albert.h.chiu@gmail.com

Insights

Angio-Seal demonstrated a lower primary failure rate compared to Perclose for antegrade common femoral artery punctures. Further research is recommended to validate these findings for closure devices in antegrade punctures.

Area of Science:

  • Vascular Surgery
  • Interventional Cardiology
  • Medical Device Technology

Background:

  • Vascular closure devices are crucial for hemostasis after peripheral angioplasty.
  • Common femoral artery (CFA) punctures can be performed using antegrade or retrograde approaches.
  • Comparing the efficacy and safety of different closure devices is essential for optimizing patient outcomes.

Purpose of the Study:

  • To compare the efficacy, primary failure rates, and complications of Angio-Seal and Perclose devices.
  • To evaluate device performance in both antegrade and retrograde common femoral artery (CFA) punctures.
  • To identify potential differences in outcomes based on puncture direction and device type.

Main Methods:

  • Retrospective single-center analysis of 123 CFA punctures in 97 patients undergoing peripheral angioplasty.
  • Patients were categorized by puncture direction (antegrade or retrograde) and closure device (Angio-Seal or Perclose).
  • Outcomes assessed included primary endpoint achievement, primary failure rates, and postprocedural complications.

Main Results:

  • No significant difference in the primary endpoint (complication-free progress to discharge) between devices or puncture directions.
  • Angio-Seal showed a significantly lower primary failure rate than Perclose in antegrade punctures (11.1% vs. 38.7%, p=0.038).
  • Antegrade punctures had a higher overall primary failure rate, but this did not lead to increased postprocedural complications.

Conclusions:

  • Angio-Seal is associated with a lower primary failure rate than Perclose for antegrade CFA punctures.
  • The study highlights the need for larger investigations to compare closure device efficacy specifically in antegrade punctures.