Related Experiment Video
Updated: Jul 4, 2026

Measurement of Healthy and Injured Triceps Surae Morphology
Published on: October 27, 2023
The detection of full thickness rotator cuff tears using ultrasound
A Al-Shawi1, R Badge, T Bunker
1Princess Elizabeth Orthopaedic Centre, Exeter, England.
Abstract:
We have examined the accuracy of 143 consecutive ultrasound scans of patients who subsequently underwent shoulder arthroscopy for rotator-cuff disease. All the scans and subsequent surgery were performed by an orthopaedic surgeon using a portable ultrasound scanner in a one-stop clinic. There were 78 full thickness tears which we confirmed by surgery or MRI. Three moderate-size tears were assessed as partial-thickness at ultrasound scan (false negative) giving a sensitivity of 96.2%. One partially torn and two intact cuffs were over-diagnosed as small full-thickness tears by ultrasound scan (false positive) giving a specificity of 95.4%. This gave a positive predictive value of 96.2% and a negative predictive value of 95.4%. Estimation of tear size was more accurate for large and massive tears at 96.5% than for moderate (88.8%) and small tears (91.6%). These results are equivalent to those obtained by several studies undertaken by experienced radiologists. We conclude that ultrasound imaging of the shoulder performed by a sufficiently-trained orthopaedic surgeon is a reliable time-saving practice to identify rotator-cuff integrity.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called a...