Avecor Trillium oxygenator versus noncoated Monolyth oxygenator: a prospective randomized controlled study
Frédéric Vanden Eynden1, Michel Carrier, Sandra Ouellet
1Department of Cardiac Surgery, Erasmus Hospital, Free University of Brussels, Belgium.
Surface coating on cardiopulmonary bypass (CPB) oxygenators did not affect blood loss or inflammation markers. However, the coated Avecor oxygenator showed a higher rate of neurological events compared to uncoated devices.
Area of Science:
- Biomedical Engineering
- Cardiovascular Surgery
- Materials Science
Background:
- Cardiopulmonary bypass (CPB) is associated with adverse effects.
- Surface coatings aim to mitigate CPB-induced harm.
- Investigating biomaterials like heparin, PEO, and sulfate/sulfonate groups for oxygenator coatings.
Purpose of the Study:
- To evaluate the impact of a coated hollow fiber membrane oxygenator on coagulation, inflammation, and clinical outcomes.
- To compare a biopassive surface oxygenator (Avecor) with a standard uncoated oxygenator (Monolyth).
Main Methods:
- Randomized trial of 200 patients undergoing valve surgery with or without coronary artery bypass grafting.
- Patients received CPB using either the coated Avecor or uncoated Monolyth oxygenator.
- Primary and secondary endpoints included biochemical markers, coagulation profiles, inflammatory mediators, blood loss, and neurological events.
Main Results:
- No significant differences were observed in biochemistry, coagulation profiles, inflammatory mediator release, or blood loss between the coated and uncoated oxygenator groups.
- Five patients in the coated oxygenator group experienced strokes, confirmed by CT scan, while no strokes occurred in the uncoated group.
Conclusions:
- The Avecor coated oxygenator demonstrates comparable performance to standard uncoated oxygenators regarding inflammation and coagulation.
- An elevated rate of neurological events was observed in patients using the Avecor oxygenator, warranting further investigation.
- Additional research is necessary to elucidate the reasons behind the increased neurological events.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...