Temporal artery temperature measurements do not detect hyperthermic marathon runners
Kevin Ronneberg1, William O Roberts, Alexander Duncan McBean
1Fairview Sports and Orthopedic Care, Wyoming, MN, USA.
Introduction:
Exertional heat stroke is a cause of collapse in marathon runners. Rectal temperature (T(rectal)) measurement is the usual method of estimating core temperature in collapsed runners, and temporal artery thermometer (TAT) measurement is untested for field use in marathon runners and other athletes. The objective of this study is to compare TAT measurement with T(rectal) measurement in collapsed marathon runners.
Methods:
TAT-500i (Exergen Corp, Wellesley, MA) temperature measurements were obtained using the manufacturer's instruction manual on 60 collapsed marathon runners who had T(rectal) measurements in the finish area medical tent during two consecutive annual races.
Results:
The TAT temperatures identified only 2 of 17 hyperthermic runners (T(rectal) > 39.4 degrees C (103 degrees F)), a sensitivity of only 0.12, and a Pearson' correlation coefficient of 0.374 (r = 0.14). Among the 17 hyperthermic runners, the correlation of T(rectal) to TAT temperatures was 0.526 (r = 0.28) with a mean +/- SD T(rectal) of 40.7 +/- 0.94 degrees C (105.1 +/- 1.7 degrees F) and a mean +/- SD TAT temperature of 37.4+/- 1.3 degrees C (99.4 +/- 2.4 degrees F). Among the 43 collapsed normothermic runners, there was no correlation between the rectal and the TAT measurements (r = -0.142, P = 0.37).
Conclusions:
These findings indicate that there is little association between the temperatures obtained by temporal artery measurement and T(rectal) measurement in collapsed marathon runners and that TAT temperature is unable to identify hyperthermic runners. Based on these findings, TAT measurement should not be used to assess core body temperature or make treatment decisions for marathon runners with potential exertional heat stroke.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...