Robotized Testing of Camera Positions to Determine Ideal Configuration for Stereo 3D Visualization of Open-Heart Surgery
Published on: August 12, 2021
Effect of reduced stereoscopic camera separation on ring placement with a surgical telerobot
Jonathan M Fishman1, Stephen R Ellis, Christopher J Hasser
1Department of Surgery, John Radcliffe Hospital, University of Oxford, Oxford, England, UK. jfishman@doctors.org.uk
Background:
A custom, stereoscopic video camera was built to study the impact of decreased camera separation on a stereoscopically viewed, visual-manual task resembling some aspects of surgery.
Materials And Methods:
Twelve naïve subjects and one of the experimenters were first trained in a ring placement task using the stereo-laparoscope and subsequently switched to the stereo-camera, which was used with differing camera separations ranging from 100% of the laparoscope's separation to a biocular view corresponding to no separation (2D).
Results:
The results suggest firstly, that stereopsis (i.e., use of 3D laparoscopes) improves surgical performance over conventional 2D laparoscopes, and secondly that camera separation may be reduced 20-35% without appreciably degrading user performance. Even a 50% reduction in separation resulted in stereoscopically supported performance far superior compared to the 2D condition.
Conclusions:
The results suggest that existing 3D laparoscopes which use 5-mm camera separation may well be significantly miniaturized without causing substantial performance degradation.