HPLC with UV or mass spectrometric detection for quantifying endogenous uracil and dihydrouracil in human plasma
Rūta Svobaite1, Isabelle Solassol, Frederic Pinguet
1Pharmacokinetic Laboratory, Faculty of Pharmacy, University Montpellier I, Montpellier, France.
Background:
We developed and compared 2 different methods for quantifying uracil (U) and dihydrouracil (UH(2)) in BSA and human plasma. Special attention was paid to the selectivity/specificity and the absence of a matrix effect. The UH(2)/U ratio is intended as a biomarker to identify patients with deficiency in 5-fluorouracil metabolism.
Methods:
We quantified U and UH(2) with 2 liquid chromatography methods after solid-phase extraction, one with UV detection (LC-UV) and the other with mass spectrometric detection (LC-MS). We selected 2 internal standards to prevent the risk of interferences. Separation was achieved with a Waters Atlantis dC18 column (LC-MS) or a Waters SymmetryShield RP18 column connected with an Atlantis dC18 (LC-UV). Mass spectrometric data were acquired in single-ion monitoring mode.
Results:
Assay imprecision in BSA solution was <15% (LC-UV) and <12% (LC-MS); in plasma, assay imprecision was <9.5% and <9.0%, respectively. Recoveries were 88.2%-110% (LC-UV) and 94.8%-107% (LC-MS). Extraction efficiencies were >or=89.0%. In BSA, the lower limits of quantification for U and UH(2) were 2.5 microg/L and 6.25 microg/L, respectively, for the LC-UV method and 2.5 microg/L and 3.1 microg/L for LC-MS. The corresponding values in plasma were 11.6 microg/L and 21.5 microg/L, and 4.1 microg/L and 12.1 microg/L.
Conclusions:
To estimate endogenous U and UH(2) concentrations and their ratio, we recommend the use of a drug-free human plasma pool in which baseline U and UH(2) concentrations have previously been measured with the standard-addition method. Our LC-MS method, which has the better test performance and is useful for measuring UH(2)/U ratios in cancer patients, is preferred when this equipment is available.