BAHA devices and magnetic resonance imaging scanners
Michael H Fritsch1, Ilka C Naumann, Kristine M Mosier
1Department of Otolaryngology-Head and Neck Surgery, Indiana University Medical Center, Indianapolis, Indiana 46202, USA. mfritsch@iupui.edu
The Bone-anchored Hearing Aid (BAHA System) fixture and abutment showed no movement in cadaver studies up to 9.4 Tesla MRI. Patients with BAHA Systems can safely undergo MRI scans with minimal image artifact.
Area of Science:
- Medical Devices
- Audiology
- Radiology
Background:
- Patients with Bone-anchored Hearing Aid (BAHA) systems may need MRI scans.
- Assessing the safety of BAHA systems in MRI environments is crucial for patient care.
Purpose of the Study:
- To evaluate the magnetic field interactions of the BAHA System fixture and abutment.
- To determine the safety and potential image artifacts of BAHA System devices during 9.4-Tesla MRI exposure.
Main Methods:
- Testing BAHA System fixtures (3 and 4 mm abutments) in Petri dishes and human cadaver temporal bones under a 9.4-T MRI magnet.
- Observing the external BAHA processor's behavior in the MRI suite.
- Reviewing MRI scans of two patients with BAHA devices at 1.5 T and 3.0 T.
Main Results:
- The BAHA System fixture and abutment did not move in cadaver temporal bone models.
- The external BAHA processor exhibited significant movement towards the magnet.
- Patient MRIs showed no adverse effects, with image quality unaffected beyond 1 cm from the BAHA post.
Conclusions:
- The bone-mounted BAHA System fixture and abutment are safe up to 9.4 T, resisting magnetic displacement.
- MRI scanning of patients with BAHA Systems is safe at field strengths up to 3.0 T and likely higher.
- Minimal artifactation around the BAHA post allows for excellent image quality at a distance.
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic Stimulation (TMS).
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Applications Of NMR In Biology
The...