A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
Reliable suction detection for patients with rotary blood pumps
David G Mason1, Andrew K Hilton, Robert F Salamonsen
1Department of Surgery, Monash University, Melbourne, Australia. dgmason62@hotmail.com
Abstract:
All rotary blood pumps (RBPs) are prone to the harmful effects of ventricular collapse or "suction events" because of over-pumping, because they are inherently preload insensitive devices, yet RBP controllers do not comprise a clinically reliable suction detector. We therefore investigated the clinical performance of seven expertly selected time domain indices of suction based on the observed positive spike induced in the RBP impeller speed waveform. Using expert panel classifications, a balanced set of 404 five-second speed snapshots of normal and suction events was created from the impeller speed 25 Hz data in 12 VentrAssist implant patients. Initially, suction index threshold levels were set differently for each patient, giving best sensitivity 95% and specificity 99%. However, analysis of paired combinations of suction indices with fixed thresholds identified one pair giving an acceptable sensitivity of 99.5% and specificity 97.5%; the low number of high speed data samples relative to the speed snapshot mean and maximum OR the largest increase in successive speed maxima. The additional precondition of RBP speed amplitude exceeding a low threshold level allows its more general application to patients with low cardiac contractility. This gives a suction detector with high clinical utility; requiring three index threshold settings only.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Suctioning the Nasopharyngeal Airway
Equipment Required