In Vitro Evaluation of an Immediate Response Starling-Like Controller for Dual Rotary Blood Pumps
Andrew F Stephens1,2, Michael C Stevens1,3,4, Shaun D Gregory1,2,5
1Innovative Cardiovascular Engineering and Technology Laboratory, Critical Care Research Group, The Prince Charles Hospital, Brisbane, QLD, Australia.
Abstract:
Rotary ventricular assist devices (VADs) are used to provide mechanical circulatory support. However, their lack of preload sensitivity in constant speed control mode (CSC) may result in ventricular suction or venous congestion. This is particularly true of biventricular support, where the native flow-balancing Starling response of both ventricles is diminished. It is possible to model the Starling response of the ventricles using cardiac output and venous return curves. With this model, we can create a Starling-like physiological controller (SLC) for VADs which can automatically balance cardiac output in the presence of perturbations to the circulation. The comparison between CSC and SLC of dual HeartWare HVADs using a mock circulation loop to simulate biventricular heart failure has been reported. Four changes in cardiovascular state were simulated to test the controller, including a 700 mL reduction in circulating fluid volume, a total loss of left and right ventricular contractility, reduction in systemic vascular resistance ( SVR) from 1300 to 600 dyne s/cm5, and an elevation in pulmonary vascular resistance ( PVR) from 100 to 300 dyne s/cm5. SLC maintained the left and right ventricular volumes between 69-214 mL and 29-182 mL, respectively, for all tests, preventing ventricular suction (ventricular volume = 0 mL) and venous congestion (atrial pressures > 20 mm Hg). Cardiac output was maintained at sufficient levels by the SLC, with systemic and pulmonary flow rates maintained above 3.14 L/min for all tests. With the CSC, left ventricular suction occurred during reductions in SVR, elevations in PVR, and reduction in circulating fluid simulations. These results demonstrate a need for a physiological control system and provide adequate in vitro validation of the immediate response of a SLC for biventricular support.
More Related Videos
Related Concept Videos
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
PID Controller
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation....
Motor Unit Stimulation
The latent period of contraction marks the onset of excitation-contraction coupling, when the action potential propagates across the sarcolemma, preparing the muscle fibers for contraction. As the fibers enter the contraction phase, the...