Related Experiment Video
Updated: Jul 3, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Computer-assisted and conventional total knee replacement: a comparative, prospective, randomised study with
J Lützner1, F Krummenauer, C Wolf
1Department of Orthopaedic Surgery, University Hospital Carl Gustav Carus, Medical Faculty, Technical University of Dresden, Fetscherstrasse 74, D-01307 Dresden, Germany. Joerg.Luetzner@uniklinikum-dresden.de
Abstract:
After obtaining informed consent, 80 patients were randomised to undergo a navigated or conventional total knee replacement. All received a cemented, unconstrained, cruciate-retaining implant with a rotating platform. Full-length standing and lateral radiographs and CT scans of the hip, knee and ankle joint were carried out five to seven days after operation. No notable differences were found between computer-assisted navigation and conventional implantation techniques as regards the rotational alignment of the femoral or tibial components. Although the deviation from the transepicondylar axis was relatively low, there was a considerable range of deviation for the tibial rotational alignment. There was no statistically significant difference regarding the occurrence pattern of outliers in mechanical malalignment but the number of outliers was reduced in the navigated group.