Abstract:
A new medical device has been designed for inducing hypothermia in kidneys, via laparoscopic deployment, without leaving residual fluid in the abdomen. It was tested in laboratory and animal trials to ascertain its suitability for achieving a target renal temperature of 15 degrees C for tissue preservation within a 30-min period. A simple theoretical model was developed to calculate cooling rates under ideal circumstances. In the laboratory, a potato was used as the model for the kidney and an average minimum temperature of 15.8 degrees C was reached in 30 min. The theoretical model calculated this as 9.9 degrees C. In the animal trials, an average minimum temperature of 11.4 degrees C was reached in 30 min, compared to the calculated temperature of 9.8 degrees C from the theoretical model. In conclusion, the new device can cool as affectively as other techniques trialled, with the added advantage that it does not release cold fluid or ice into the abdomen.
More Related Videos
03:13Technical Refinement of a Bilateral Renal Ischemia-Reperfusion Mouse Model for Acute Kidney Injury Research
Published on: November 3, 2023
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Related Concept Videos
Kidney Transplant II: Surgical Procedure
Urinary Tract Calculi VI: Surgical Management
Extracorporeal Removal of Drugs: Continuous Renal Replacement Therapy