Related Experiment Video
Updated: Jul 3, 2026

Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Dosimetry of the microSelectron-HDR Ir-192 source using PRESAGE and optical CT
P Wai1, J Adamovics, N Krstajic
1Department of Physics, School of Electronics & Physical Sciences, University of Surrey, Guildford, Surrey, UK. P.Wai@surrey.ac.uk
Abstract:
Optical CT, using a solid polyurethane (PRESAGE) radiochromic dosimeter, has been used to evaluate dose distributions produced by the microSelectron-HDR Ir-192 source. The anisotropy functions obtained through optical CT are in good agreement with Monte Carlo and previously published results especially at polar angle above 20 degrees. The results indicated an evident potential for using solid polymer dosimetry as an accurate method for 3-D dosimetry, although refinements to the existing methods are necessary before the technique can be used clinically.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET