Maki Ichinose1, Hiroaki Sakai, Ikuo Miyazaki
1Department of Anesthesiology, Juntendo University Shizuoka Hospital, 1129 Nagaoka, Izunokuni, Shizuoka 410-2295, Japan.
Abstract:
This case report describes the difficult respiratory management of an esophageal cancer patient with acute respiratory distress syndrome (ARDS) and systemic inflammatory response syndrome (SIRS) caused by a postoperative tracheogastric roll fistula. A single-lumen tracheal tube could not seal the fistula, and therefore a double-lumen tracheal tube (DLT) for the left side was used. Although the proximal cuff of the DLT failed to seal the fistula, independent lung ventilation (ILV) improved blood gas levels. During right thoracotomy, the left lung was ventilated conventionally with 5 cmH2O positive end-expiratory pressure (PEEP), and in addition, high-frequency oscillation ventilation (HFOV) to the right lung was employed. This combination allowed the maintenance of adequate oxygenation, and the HFOV to the right lung decreased the PaCO2 level during surgery without interruption of the surgical field. These techniques provided the opportunity to successfully remove a necrotic gastric roll and achieve closure of the fistula using an intercostal muscle flap. This report documents and discusses the difficulty of performing appropriate anesthetic management of a patient with these complex complications after esophageal surgery.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...