Related Experiment Video
Updated: Jul 2, 2026

Quantifying Microorganisms at Low Concentrations Using Digital Holographic Microscopy (DHM)
Published on: November 1, 2017
Retrospective determination of the contamination in the HML's counting chambers
Gary H Kramer1, Barry Hauck, Kevin Capello
1Human Monitoring Laboratory, Radiation Surveillance and Health Assessment Division, Radiation Protection Bureau, 775 Brookfield Road, PL6302D1, Ottawa, Ontario, Canada. gary_h_kramer@hc-sc.gc.ca
Abstract:
The original documentation surrounding the purchase of the Human Monitoring Laboratory's (HML) counting chambers clearly showed that the steel contained low levels of radioactivity, presumably as a result of A-bomb fallout or perhaps to the inadvertent mixing of radioactive sources with scrap steel. Monte Carlo simulations have been combined with experimental measurements to estimate the level of contamination in the steel of the HML's whole body counting chamber. A 24-h empty chamber background count showed the presence of 137Cs and 60Co. The estimated activity of 137Cs in the 51 tons of steel was 2.7 kBq in 2007 (51.3 microBq g(-1) steel) which would have been 8 kBq at the time of manufacture. The 60Co that was found in the background spectrum is postulated to be contained in the bed-frame. The estimated amount in 2007 was 5 Bq and its origin is likely to be contaminated scrap metal entering the steel production cycle sometime in the past. The estimated activities are 10 to 25 times higher than the estimated minimum detectable activity for this measurement. These amounts have no impact on the usefulness of the whole body counter.