Identification of a previously undetected source of systematic error in capillary viscometry measurements
G D Wedlake1, J H Vera, G A Ratcliff
1Department of Chemical Engineering, University of Alberta, Edmonton, Alberta T6G 2G6 CanadaDepartment of Chemical Engineering, McGill University, Montreal, Quebec H3A 2A7 Canada.
Abstract:
A new analysis of the factors affecting the driving head within a capillary viscometer has identified a previously undetected source of error due to surface tension. In contrast to the well-known capillary rise correction to the effective driving head, this additional effect requires that a correction be made to the apparent head of liquid. Calculations show that this source of error may be responsible for deviations in experimental results as large as 1%. A new calibration equation which compensates for both surface tension effects is recommended.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Capillary Electrophoresis: Instrumentation
Random and Systematic Errors
Data Validation
Key parameters for method validation include:
Uncertainty in Measurement: Accuracy and Precision
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...