Related Experiment Videos
Induced effects in ionization chamber cables by photon and electron irradiation
1Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, São Paulo, Brazil.
Medical Physics
|May 1, 1991
Summary
Irradiation affects ionization chamber cables and connectors. Correction factors were determined based on radiation energy, cable length, and absorbed dose for accurate measurements.
Area of Science:
- Medical Physics
- Radiation Detection and Measurement
Background:
- Ionization chambers are crucial for radiation dosimetry.
- Cables and connectors can be susceptible to radiation-induced effects, impacting measurement accuracy.
- Understanding these effects is vital for reliable radiation monitoring.
Purpose of the Study:
- To investigate the impact of photon and electron irradiation on ionization chamber cables and connectors.
- To quantify the radiation-induced leakage current and its influence on chamber response.
- To develop correction factors for irradiation effects.
Main Methods:
- Experimental investigation under varied conditions.
- Measurement of ionization chamber response using a 90Sr standard check source.
- Measurement of radiation-induced leakage current.
Main Results:
- Photon and electron irradiation alter the performance of cables and connectors.
- Radiation-induced leakage currents were observed and measured.
- Correction factors were successfully derived, dependent on radiation energy, cable length, and absorbed dose.
Conclusions:
- Irradiation effects on ionization chamber accessories are significant and quantifiable.
- The derived correction factors enable compensation for these effects.
- Accurate radiation measurements can be maintained by applying these correction factors.