Related Experiment Video
Updated: Jul 2, 2026

A Porcine Model of Acute Respiratory Failure with a Continuous Infusion of Oleic Acid
Published on: March 8, 2024
A pilot study of neonatal and pediatric esophageal pulse oximetry
Panayiotis A Kyriacou1, Deric P Jones, Richard M Langford
1School of Engineering and Mathematical Sciences, City University, London, EC1V 0HB, UK. p.kyriacou@city.ac.uk
Background:
In this pilot study we explored the suitability of the esophagus as a new measuring site for blood oxygen saturation (Spo(2)) in neonates.
Methods:
A new miniaturized esophageal pulse oximeter has been developed. Five patients (one child and four neonates) were studied.
Results:
Spo(2) values were obtained in the esophagus of all patients. A Bland and Altman plot of the difference between Spo(2) values from the esophageal pulse oximeter and a commercial toe pulse oximeter against their mean showed that the bias and the limits of agreement between the two pulse oximeters were +0.3% and +1.7% to -1.0%, respectively.
Conclusions:
This study suggests that the esophagus can be used as an alternative site for monitoring blood oxygen saturation in children and neonates.