Relationship between density variability and imaging volume size in cone-beam computerized tomographic scanning of
Akitoshi Katsumata1, Akiko Hirukawa, Shinji Okumura
1Department of Oral Radiology, Asahi University School of Dentistry, Gifu, Japan. kawamata@dent.asahi-u.ac.jp
Objective:
In limited-volume cone-beam computerized tomography (CBCT) imaging, projection data discontinuity caused by maxillofacial hard tissue structures outside the reconstructed volume are reported to affect the density value of the hard and soft tissue structures within the volume. The intensity of this effect is purported to be related to the size of the imaging volume. The aim of this study was to characterize the effect of the size of the scanned volume on density values in vitro.
Study Design:
Test objects were positioned in a custom phantom in the following 4 patterns: bimandible and vertebrae, bimandible, left mandible and vertebrae, and left mandible. We used a newly developed flat panel detector CBCT system (Alphard Vega; Asahi Roentgen, Kyoto, Japan) to acquire scans of the left molar region using cylindrical volumes of approximately 5, 10, 15, and 20 cm in diameter and height. The density values of the mandible and the adjacent soft tissue regions were analyzed.
Results:
Highest density variability was observed in the smallest-volume (5 cm) scans. Density variability increased when more objects were included outside the area being imaged. Fewer effects were noted in CBCT scans of larger (10, 15, and 20 cm) volumes.
Conclusion:
Larger-volume CBCT scans may yield more consistent density values. Smaller CBCT volumes have the advantages of better image resolution and lower radiation doses. The optimization of the image characteristics is maximized by careful consideration of the purpose of the CBCT examination.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...