Related Experiment Video
Updated: Jul 2, 2026

In Vitro 3D Cell-Cultured Arterial Models for Studying Vascular Drug Targeting Under Flow
Published on: March 14, 2021
Modeling of human colonic blood flow for a novel artificial anal sphincter system
Peng Zan1, Guo-zheng Yan, Hua Liu
1School of Electronics, Information and Electrical Engineering, Shanghai Jiao Tong University, Shanghai 200240, China. zanpeng@sjtu.edu.cn
Abstract:
A novel artificial anal sphincter system has been developed to simulate the normal physiology of the human anorectum. With the goal of engineering a safe and reliable device, the model of human colonic blood flow has been built and the relationship between the colonic blood flow rate and the operating occlusion pressure of the anorectum is achieved. The tissue ischemia is analyzed based on constitutive relations for human anorectum. The results suggest that at the planned operating occlusion pressure of less than 4 kPa the artificial anal sphincter should not risk the vascularity of the human colon.