Feasibility of volume-of-interest (VOI) scanning technique in cone beam breast CT--a preliminary study
Lingyun Chen1, Chris C Shaw, Mustafa C Altunbas
1Department of Imaging Physics, University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030-4009, USA. lingyun.chen@di.mdacc.tmc.edu
Abstract:
This work is to demonstrate that high quality cone beam CT images can be generated for a volume of interest (VOI) and to investigate the exposure reduction effect, dose saving, and scatter reduction with the VOI scanning technique. The VOI scanning technique involves inserting a filtering mask between the x-ray source and the breast during image acquisition. The mask has an opening to allow full x-ray exposure to be delivered to a preselected VOI and a lower, filtered exposure to the region outside the VOI. To investigate the effects of increased noise due to reduced exposure outside the VOI on the reconstructed VOI image, we directly extracted the projection data inside the VOI from the full-field projection data and added additional data to the projection outside the VOI to simulate the relative noise increase due to reduced exposure. The nonuniform reference images were simulated in an identical manner to normalize the projection images and measure the x-ray attenuation factor for the object. Regular Feldkamp-Davis-Kress filtered backprojection algorithm was used to reconstruct the 3D images. The noise level inside the VOI was evaluated and compared with that of the full-field higher exposure image. Calcifications phantom and low contrast phantom were imaged. Dose reduction was investigated by estimating the dose distribution in a cylindrical water phantom using Monte Carlo simulation based Geant4 package. Scatter reduction at the detector input was also studied. Our results show that with the exposure level reduced by the VOI mask, the dose levels were significantly reduced both inside and outside the VOI without compromising the accuracy of image reconstruction, allowing for the VOI to be imaged with more clarity and helping to reduce the breast dose. The contrast-to-noise ratio inside the VOI was improved. The VOI images were not adversely affected by noisier projection data outside the VOI. Scatter intensities at the detector input were also shown to decrease significantly both inside and outside the VOI in the projection images, indicating potential improvement of image quality inside the VOI and contribution to dose reduction both inside and outside the VOI.
More Related Videos
06:03Integrating Augmented Reality Tools in Breast Cancer Related Lymphedema Prognostication and Diagnosis
Published on: February 6, 2020
12:23Multi-modal Imaging of Angiogenesis in a Nude Rat Model of Breast Cancer Bone Metastasis Using Magnetic Resonance Imaging, Volumetric Computed Tomography and Ultrasound
Published on: August 14, 2012
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...