Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
Identification of coagulase-negative Staphylococci by using the BD phoenix system in the low-inoculum mode
Gioconda Brigante1, Maria Grazia Menozzi, Beatrice Pini
1Department of Microbiology, University of Catania, Via Androne 81, 95124 Catania, Italy.
Abstract:
The new "low-inoculum" mode of the Phoenix system was evaluated to identify clinical coagulase-negative staphylococci. API ID32 Staph panels were used as comparators, and discrepancies were resolved by 16S rRNA and tuf gene analysis. The system correctly identified 90.5% of isolates, with a mean time of 10.2 h. Accuracy was satisfactory for Staphylococcus epidermidis, S. saprophyticus, and S. haemolyticus.