Related Experiment Video
Updated: Jun 30, 2026

Alternating Magnetic Field-Responsive Hybrid Gelatin Microgels for Controlled Drug Release
Published on: February 13, 2016
Reduced acute inflammatory responses to microgel conformal coatings
Amanda W Bridges1, Neetu Singh, Kellie L Burns
1Petit Institute for Bioengineering and Bioscience, 315 Ferst Drive, 2314 IBB, Atlanta, GA 30332-0363, USA.
Abstract:
Implantation of synthetic materials into the body elicits inflammatory host responses that limit medical device integration and biological performance. This inflammatory cascade involves protein adsorption, leukocyte recruitment and activation, cytokine release, and fibrous encapsulation of the implant. We present a coating strategy based on thin films of poly(N-isopropylacrylamide) hydrogel microparticles (i.e. microgels) cross-linked with poly(ethylene glycol) diacrylate. These particles were grafted onto a clinically relevant polymeric material to generate conformal coatings that significantly reduced in vitro fibrinogen adsorption and primary human monocyte/macrophage adhesion and spreading. These microgel coatings also reduced leukocyte adhesion and expression of pro-inflammatory cytokines (TNF-alpha, IL-1beta, MCP-1) in response to materials implanted acutely in the murine intraperitoneal space. These microgel coatings can be applied to biomedical implants as a protective coating to attenuate biofouling, leukocyte adhesion and activation, and adverse host responses for biomedical and biotechnological applications.