Spectrally average conversion coefficients for air kerma to ambient dose equivalent for clinical linear accelerator
Marco A Frota1, Verginia R Crispim, Ademir X Silva
1Laboratório de Radiobiologia e Radiometria, Departamento de Biologia Geral, Universidade Federal Fluminense, Outeiro de São João Batista, s/n, Caixa Postal 100.436-24001-970 Niterói, RJ, Brazil.
Abstract:
This work aims to calculate the conversion coefficients from air kerma to ambient dose equivalent, H*(10)/K(air) for photon beams produced by linear accelerators, such as the Clinac-4, Clinac-6, Clinac-18 and Clinac-2500, after transmission through primary barriers of radiotherapy treatment rooms. Concrete walls of thickness 1.0, 1.5 and 2.0m were irradiated with 30cmx30cm primary beam spectra. The transmitted spectra were calculated to obtain the conversion coefficients for beams found in radiotherapy services. The calculations were done using the MCNP-4B Monte Carlo code. The results indicate the need to use a factor of about 1.20 to obtain the ambient dose equivalent for radiation surveys near primary barriers using instruments calibrated in air kerma.